Table of Contents
- What Is the EU MDR?
- Who Must Comply With the EU MDR?
- EU MDR Device Classification
- Which Substances Does the EU MDR Restrict?
- EU MDR Transition Deadlines
- EU MDR vs MDD: What Changed?
- EU MDR Compliance Requirements
- The role of notified bodies
- UDI and EUDAMED
- EU MDR vs IVDR
- Penalties and Risks of Non-Compliance
- How Regilient Helps With MDR Substance Data
The EU Medical Device Regulation (EU) 2017/745, usually called the EU MDR, sets the rules for placing medical devices on the EU market. It has applied since 26 May 2021 and replaced the Medical Devices Directive (93/42/EEC) and the Active Implantable Medical Devices Directive (90/385/EEC).
EU MDR in brief:
- It is a regulation, so it applies directly in every EU member state without national transposition.
- Devices are sorted into four risk classes (I, IIa, IIb and III), and most classes need a notified body.
- Manufacturers need a quality management system, technical documentation, clinical evidence, post-market surveillance and a Person Responsible for Regulatory Compliance (PRRC).
- Devices carry a Unique Device Identification (UDI) and are registered in the EUDAMED database.
- Legacy devices certified under the old directives can stay on the market only until 31 December 2027 or 31 December 2028, depending on class, and only if conditions are met.
Last reviewed: 11 November 2026. This guide is informational and is not legal or regulatory advice. Confirm dates and obligations against the official text and your notified body.
What Is the EU MDR?
The EU MDR (Regulation (EU) 2017/745) governs the design, manufacture, placing on the market and post-market monitoring of medical devices. Its aim is to improve patient safety, clinical evidence and traceability across the full device lifecycle, not just at the point of CE marking.
Compared with the directives it replaced, the MDR sets stricter requirements for clinical evaluation, gives notified bodies tougher oversight, and makes manufacturers responsible for ongoing performance data. It also brings in some products without a medical purpose, listed in Annex XVI, such as certain contact lenses and aesthetic equipment.
Who Must Comply With the EU MDR?
- Manufacturers, whether based inside or outside the EU.
- Authorised representatives, which non-EU manufacturers must appoint to sell in the EU.
- Importers and distributors, who must verify CE marking, labelling and documentation before supplying a device.
- OEMs and private labellers, who take on manufacturer obligations when they place a device on the market under their own name.
EU MDR Device Classification
Classification depends on the intended purpose, how invasive the device is, how long it is used and which part of the body it affects. The classification rules sit in Annex VIII. Many software products and some devices that were Class I under the MDD now fall into a higher class.
Class | Risk level | Typical examples | Conformity route |
|---|---|---|---|
Class I | Low | Gauze, spectacle frames, stethoscopes | Manufacturer self-declaration (a notified body is needed for sterile, measuring and reusable surgical versions: Is, Im and Ir) |
Class IIa | Medium | Hearing aids, short-term catheters | Notified body |
Class IIb | Medium to high | Ventilators, long-term implantable devices | Notified body |
Class III | High | Pacemakers, prosthetic heart valves | Notified body, with the most intensive clinical review |
Which Substances Does the EU MDR Restrict?
Annex I, Section 10.4 deals with substances. It applies to devices, or parts and materials of devices, that are invasive and in direct contact with the body, or that administer, transport or store medicines or body fluids.
For those devices, the following substances may be present above 0.1% weight by weight (w/w) only where the manufacturer can justify it:
- carcinogenic or mutagenic substances of category 1A or 1B
- substances toxic to reproduction of category 1A or 1B
- substances with endocrine-disrupting properties for which there is scientific evidence of probable serious effects on human health, identified under Article 59 of REACH or by Commission delegated acts
A justification typically covers the analysis of alternatives, the benefit-risk assessment and the residual risk. Where such a substance is present above the threshold, the device or its packaging must be labelled accordingly. Phthalates have specific attention: the SCHEER guidelines on phthalates in medical devices set out how to assess their presence and justify their use.
In practice this means manufacturers need substance-level data from suppliers. Start with the REACH SVHC Candidate List, because the endocrine-disruptor identification route links to REACH, and see our guide to PFAS in medical devices.
EU MDR Transition Deadlines
Regulation (EU) 2023/607 extended the transition for devices that hold valid certificates or declarations of conformity issued under the old directives. It also removed the earlier sell-off deadline, so devices lawfully placed on the market can remain available.
Device group | End of transition |
|---|---|
Class III devices and Class IIb implantable devices (except well-established technologies such as sutures, staples, screws and plates) | 31 December 2027 |
Other Class IIb, Class IIa, sterile or measuring Class I devices, and devices that need a notified body for the first time under the MDR | 31 December 2028 |
Class III custom-made implantable devices | 26 May 2026 |
The extension is conditional. Manufacturers must keep a valid legacy certificate, make no significant changes to the design or intended purpose, have a quality management system in place, and have lodged an application with a notified body and signed a written agreement with it within the set dates. Check the conditions in Article 120 of the MDR as amended before relying on an extension.
The European Commission has also proposed further simplification of the MDR and IVDR. Treat any change as pending until it is adopted and published in the Official Journal.
EU MDR vs MDD: What Changed?
Area | MDD (93/42/EEC) | MDR (EU) 2017/745 |
|---|---|---|
Legal form | Directive, transposed nationally | Regulation, directly applicable |
Focus | Mainly pre-market conformity | Full lifecycle, from design to post-market surveillance |
Traceability | No UDI system | UDI for most devices |
Clinical evidence | Lighter requirements | Stronger clinical evaluation and follow-up |
Notified bodies | Less scrutiny | Stricter designation and oversight |
Database | Eudamed not in use | EUDAMED for actors, devices, certificates and vigilance |
EU MDR Compliance Requirements
A compliant programme usually covers the following:
- Quality management system. The MDR requires a QMS proportionate to the risk class. Many manufacturers meet this through ISO 13485.
- Technical documentation. Annexes II and III set out what the file must contain, including the device description, risk management and verification and validation evidence.
- General Safety and Performance Requirements (GSPR). Annex I requirements must be mapped to evidence in a conformity matrix.
- Clinical evaluation. Performance and safety claims need clinical data, and a plan for post-market clinical follow-up where applicable.
- Post-market surveillance. This includes a PMS plan, vigilance reporting and, for higher classes, Periodic Safety Update Reports (PSUR).
- Person Responsible for Regulatory Compliance (PRRC). Manufacturers must have this role in place.
- Labelling and instructions for use. Labels must show the UDI and the required symbols and substance information, in the languages of the countries of sale.
- Declaration of Conformity and CE marking. The declaration must reference the MDR, and the CE mark can only be applied after the conformity assessment route is complete.
The role of notified bodies
Notified bodies assess technical documentation, clinical data and the quality system for devices above Class I. Capacity has been tight since the MDR started, so allow plenty of time and agree timelines with your notified body early.
UDI and EUDAMED
The UDI system identifies each device model and production batch. UDI data is submitted to EUDAMED, the European database for actor registration, device registration, certificates, market surveillance, vigilance and clinical investigations.
EUDAMED is being switched on module by module. According to the Commission's notice, the first four modules (actor registration, UDI and device registration, notified bodies and certificates, and market surveillance) became mandatory from 28 May 2026, while the vigilance and clinical investigation modules follow later. Confirm the current position on the European Commission's EUDAMED page.
EU MDR vs IVDR
The In Vitro Diagnostic Regulation (EU) 2017/746 is the companion regulation for diagnostic devices used on samples taken from the body. A device falls under one or the other, not both. IVDR classifies devices in classes A to D and has its own timelines. See our guides to IVDR vs IVDD and EU IVDR compliance. For a side-by-side view, read MDR and IVDR compliance requirements.
Penalties and Risks of Non-Compliance
The MDR leaves penalties to member states, so they vary by country. Practical consequences include loss of CE marking validity, withdrawal or recall of devices, blocked shipments, loss of notified body certification and reputational damage with hospitals and distributors. Authorities can also request documentation at any time, so keep the technical file and supplier evidence audit-ready.
How Regilient Helps With MDR Substance Data
Regilient helps medical device manufacturers collect and manage material and substance data from suppliers, including declarations against SVHC thresholds, Full Material Disclosure (FMD) workflows and an audit trail with version control for declarations. This supports the substance evidence the MDR requires, alongside your REACH and RoHS obligations.